LC-MS/MS cutoffs: Morphine & Codeine = 50 ng/mL (urine), 1 ng/mL (oral fluid). Any result below these thresholds is clinically and legally Negative. Results are truncated (not rounded): 19.9 becomes 19. The 1% Rule requires the parent compound to be at or above 50 ng/mL before it can be applied to explain secondary opioids.
| Creatinine (mg/dL) | Interpretation |
|---|---|
| < 2.0 | Physiologically Impossible: not human urine |
| 2.0 – 5.0 | Highly Improbable: suggests direct water addition |
| 5.0 – 20.0 | Dilute: excessive fluid intake or water addition |
| 2.0 – 20.0 with SG > 1.020 | Probable Non-Urine: possible substitution (e.g., juice) |
| > 20.0 | Normal range |
| SG Value | Interpretation |
|---|---|
| < 1.003 | Dilute / Tampered |
| 1.003 – 1.020 | Normal |
| > 1.020 (with Cr 2–20) | Suspicious: possible non-urine substitution |
| pH Range | Interpretation |
|---|---|
| 4.5 – 8.0 | Normal physiological range |
| Up to 9.5 | Clinical acceptance limit |
| < 4.5 or > 8.0 (SAMHSA forensic limit: 9.1) | Tampered |
- Bleach (sodium hypochlorite): oxidizing agent; degrades drug metabolites; detected via oxidant screen and abnormal odor
- Vinegar / ammonia: shifts pH to extremes; detected by pH testing
- Detergents / soaps: increase viscosity; cause foamy or scented urine
- Commercial products (Urine Luck, CleanX): interfere with immunoassays
- Powdered or synthetic urine: abnormal SG/creatinine pairing, often gelatinous texture
SVT evaluates four markers: creatinine, specific gravity, oxidant activity, and pH. Each is reported Normal or Abnormal against established reference ranges. Abnormal results have multiple causes: excessive hydration, medical conditions, medications, and tampering. No single marker confirms tampering on its own.
| Marker | Normal Range | What Can Cause an Abnormal Result |
|---|---|---|
| Creatinine | > 20 mg/dL | Excessive fluid intake, diuretics, water addition, or renal disease. Below 2 mg/dL is not physiologically possible (not human urine). |
| Specific Gravity | 1.003 to 1.035 | Low with low creatinine: dilution or water addition. Discordant high SG with low creatinine: possible salt addition or substitution. |
| Oxidant Activity | < 200 mcg/mL | Adulterants (nitrites, PCC, bleach, H2O2). Benign causes: UTI, dietary nitrites, blood. Colorimetric interference from phenazopyridine (Pyridium) or rifampin: orange-red urine absorbs at assay wavelengths, producing false or elevated oxidant readings. |
| pH | 4.5 to 9.5 | Acidic adulterants push below 4.5. Alkaline adulterants push above 9.5. pH can drift to 9.0 to 9.5 from storage or transport delay alone. |
Dilution vs. Substitution Pattern:
| Creatinine (mg/dL) | Specific Gravity | Interpretation |
|---|---|---|
| 5.0 to < 20 | < 1.003 | Dilute: over-hydration or water added |
| 2.0 to < 5.0 | < 1.003 | Highly improbable; probable water addition |
| >= 2.0 and < 20 | > 1.020 | Discordant: probably not urine; possible salt addition |
| < 2.0 | Any | Not physiologically possible: specimen is not human urine |
Medications Causing Colorimetric Interference (Oxidant Assay):
| Medication | Urine Color | Implication |
|---|---|---|
| Phenazopyridine (Pyridium, Azo) | Bright orange to orange-red | High risk of false/elevated oxidant result; not adulteration |
| Rifampin (Rifadin) | Orange-red | Colorimetric interference at therapeutic doses (TB, MAC regimens) |
| Riboflavin (vitamin B2) | Bright yellow | Rarely clinically significant for validity testing |
| Nitrofurantoin (Macrobid) | Dark yellow to brown | Low clinical significance for oxidant testing |
* Reference ranges reflect clinical laboratory thresholds used for patient care settings. See companion FAQ for federal workplace testing cutoffs.
[Gather all four validity marker results before responding if possible.]
"Let me walk through each marker and then interpret the pattern as a whole."
[Start with creatinine and SG:]
"If both are low, that is a dilute pattern from over-hydration or water addition. If creatinine is below 2 mg/dL or does not match the SG, the specimen may not be human urine."
[If oxidant is flagged:]
"Near 200 mcg/mL can reflect a UTI, dietary sources, or blood. At or above 500 mcg/mL is more consistent with an adulterant. Is the patient on phenazopyridine or rifampin? Both cause orange-red urine that interferes with the colorimetric oxidant assay."
[If pH is flagged:]
"pH can drift to 9.0 or higher from storage or transport issues alone. What was the collection temperature and how quickly was it processed?"
[Closing:]
"Read these markers together. One abnormal marker without a supporting pattern is not a basis for a tampering conclusion. If discordant, recommend observed re-collection."
Federal workplace testing (DOT, HHS-mandated programs) uses codified cutoffs that differ from clinical laboratory reference ranges. A clinically abnormal marker is not automatically adulterated or substituted under federal rules.
| Classification | Marker(s) | Federal Cutoff (HHS/SAMHSA) |
|---|---|---|
| Substituted | Creatinine + SG | Creatinine < 2 mg/dL AND SG < 1.0010 or >= 1.0200 (both required) |
| Dilute | Creatinine + SG | Creatinine >= 2 to < 20 mg/dL AND SG > 1.0010 to < 1.0030 |
| Adulterated | pH | < 4.0 or >= 11.0 |
| Adulterated | Nitrite | >= 500 mcg/mL |
| Adulterated | Chromium (VI) | >= 50 mcg/mL |
| Adulterated | Oxidant (halogen, glutaraldehyde, pyridine, surfactant) | Present at or above confirmatory test LOQ |
| Invalid | pH or nitrite (intermediate) | pH 4.0 to < 4.5 or 9.0 to < 11.0; nitrite >= 200 to < 500 mcg/mL |
Key differences from clinical ranges: Federal nitrite adulteration threshold is >= 500 mcg/mL (clinical labs may flag at >= 200 mcg/mL). A nitrite of 250 mcg/mL is clinically abnormal but federally invalid, not adulterated. Federal pH adulteration threshold is < 4.0 or >= 11.0 (clinical labs may flag outside 4.5 to 9.5). These are not interchangeable.
* Per SAMHSA MRO Guidance Manual and 49 CFR Part 40. Applies to DOT-regulated and other federally mandated testing programs only.
[MRO or drug court coordinator asks whether a result meets federal adulterated/substituted criteria.]
"First, is this a DOT-regulated test or another federally mandated program, or is it a clinical test?"
[If federal:]
"The clinical laboratory may report a marker as abnormal based on its reference ranges, but that does not automatically meet the federal threshold. Let me walk through the specific value and which federal classification, if any, applies."
[If clinical:]
"This is a clinical test, so the federal forensic cutoffs do not directly apply. We interpret against the laboratory's reference ranges and the overall validity pattern."
| Analyte | EIA Cutoff | LC-MS/MS Urine | LC-MS/MS Oral Fluid | Detection Window |
|---|---|---|---|---|
| Codeine | 300 ng/mL | 50 ng/mL | 1 ng/mL | 1–2 days |
| Morphine | 300 ng/mL | 50 ng/mL | 1 ng/mL | 1–2 days (IR); 2–3 (ER) |
Opioid Process Impurities: prescription opioids contain unavoidable manufacturing byproducts that are themselves opioids. At high doses, these impurities can be detected at low levels, raising concern about additional drug use. They are not metabolites. This concept is informally called the "1% Rule" (named because impurities are generally present at less than 1% of the parent drug concentration). The rule has one hard requirement: it cannot be applied if the parent compound is not confirmed at or above the laboratory's cutoff (typically 50 ng/mL). Without a confirmed parent, the secondary opioid must be evaluated as independent ingestion.
| Prescribed Drug | Potential Impurity | Allowable % | Typically Observed % |
|---|---|---|---|
| Codeine | Morphine | 0.15% | 0.01–0.1% |
| Morphine | Codeine | 0.5% | 0.01–0.05% |
| Hydrocodone | Codeine | 0.15% | 0–0.1% |
| Hydromorphone | Morphine, Hydrocodone | 0.15%, 0.1% | 0–0.025% |
| Oxycodone | Hydrocodone | 1.0% | 0.02–0.12% |
| Oxymorphone | Oxycodone, Hydromorphone | 0.5%, 0.15% | 0.05–0.4%, 0.03–0.1% |
* Cutoff values reflect this laboratory's thresholds. Other laboratories may use different cutoffs.
Critical Distinction: Impurity vs. Metabolite
Hydrocodone (after codeine) and hydromorphone (after morphine) are actually minor metabolites at approximately 5% of the parent drug concentration, not process impurities. This is commonly misattributed to impurities. If hydromorphone or hydrocodone appear at roughly 5% of their parent, that is metabolic, not impurity-driven.
Interpretation Criteria: a process impurity is more likely when all three conditions are present:
- The secondary opioid is not a prescribed medication
- Its urine concentration is low (generally below 100 ng/mL)
- The prescribed parent opioid is present at a high concentration (generally above 50,000 ng/mL)
Limitations
- No quantity thresholds or ratios can absolutely determine the source of an opioid
- Impurity percent varies between lots and formulations; impurities are not always detected even at elevated parent concentrations
- Interpretation must include clinical observation and professional judgment
6-AM is the unique marker for heroin use, most likely detectable within the first 24 hours. After that window, only morphine remains, which is non-specific.
In oral fluid: This is within expected range. In a study of 84,148 oral fluid specimens, 62.2% of methadone-positive results had no detectable EDDP. This is biologically normal in oral fluid and should not be interpreted as deception.
| Product | Route | Approx OF Positivity |
|---|---|---|
| Suboxone (sublingual) | Direct oral exposure | ~86% |
| Sublocade | SubQ depot injection | ~47% |
| Brixadi | SubQ depot injection | Comparable to or lower than Sublocade |
| Medication | Formulations | Brand Names | Testing Notes |
|---|---|---|---|
| Buprenorphine | SL tablets, SL films, ER injection, implant | Subutex, Suboxone, Sublocade, Brixadi, Belbuca, Zubsolv | Buprenorphine + norbuprenorphine detected; may cause weak opiate EIA cross-reactivity |
| Methadone | Oral solution, tablets, dispersible tablets | Methadose | Requires specific methadone assay; long half-life |
| Naltrexone | Oral tablets, ER monthly injection | Vivitrol, Revia | Not routinely detected on standard UDT panels |
| Naloxone | Nasal spray, IV/IM | Narcan, Kloxxado | May appear on some panels; structurally similar to morphine on EIA |
Chiral Analysis (d/l isomer testing):
| % d-isomer (mAMP) | Interpretation | Notes |
|---|---|---|
| 20% or more d-mAMP | Positive for d-isomer | Illicit mAMP, Desoxyn (Rx), or manufacturing impurity from AMP product. d-Isomer alone does NOT rule out impurity. |
| Less than 20% d-mAMP | Positive for l-isomer | OTC l-desoxyephedrine (generic vapor inhalers) or selegiline metabolism |
| Any result above 10,000 ng/mL | Strongly illicit | Effectively excludes OTC, selegiline, and manufacturing impurity |
Manufacturing Impurity Criteria (only applies when high prescription AMP is also present):
| Criterion | Impurity Pattern | Illicit Use Pattern |
|---|---|---|
| AMP result | High (prescribed; 5,000+ ng/mL) | Absent or inconsistent with prescription |
| mAMP result | Low; near or slightly above cutoff | Moderate to high; often 1,000+ ng/mL |
| mAMP:AMP ratio | Below 0.5% | Typically well above 0.5% |
| d/l isomeric correlation | mAMP and AMP d/l ratios closely match | May diverge |
| Other illicits | Absent | Often co-detected |
| Test Type | Methylphenidate Cross-Reactivity? | Clinical Action |
|---|---|---|
| POC / EMIT immunoassay | Very low with modern assays | Confirm by LC-MS/MS if unexpected positive obtained |
| LC-MS/MS (confirmatory) | None. Analyte-specific. | Confirmed positive is not a methylphenidate artifact. Assess for separate drug use. If AMP also present, see TP-MAMP-V01. |
The only primary data comes from Cone et al. (1996), a letter-level report measuring cocaine in seminal plasma of drug users. Peak seminal plasma cocaine was approximately 10 mcg/g in chronic heavy users. A typical ejaculate is 2–3 grams, yielding an estimated maximum of 20–30 mcg cocaine per encounter. The authors extrapolated approximately 10,000 precisely timed sexual encounters before a non-using partner might approach a positive drug test. No published study has documented a confirmed positive attributable to this pathway.
A separate scenario is direct urine specimen contamination (if semen is deposited shortly before voiding and collection is not done cleanly) shortly before voiding and collection is not done cleanly. If suspected, prostate-specific antigen (PSA) testing can serve as a seminal marker.
| Parameter | Value / Notes |
|---|---|
| Peak cocaine in seminal plasma | ~10 mcg/g (Cone 1996, letter-level; limited sample) |
| Estimated cocaine per ejaculate | ~20–30 mcg maximum |
| BE needed for 50 ng/mL positive (300 mL void) | ~15 mcg minimum, before distribution and clearance |
| Encounters to approach a positive (est.) | ~10,000 (author extrapolation, not empirically validated) |
| Published confirmed positives via semen | None identified in peer-reviewed literature to date |
* Cutoff values reflect this laboratory's thresholds. Other laboratories may use different cutoffs. Confirm with the reporting lab before making clinical decisions.
"Thank you for calling. I understand you have a question about a positive cocaine result in a patient claiming exposure through her partner."
[Listen for the reported concentration and collection circumstances before responding.]
"Based on the available evidence, semen-mediated cocaine transfer is not expected to produce a positive urine result at standard cutoffs. The cocaine concentrations documented in semen are too low, and the volume per encounter too small, to generate detectable benzoylecgonine in a partner's urine under typical conditions."
"That said, the published literature in this area is limited, so I would not present this as completely settled. No published study has documented a confirmed positive from this pathway."
[If caller asks about concentration:] "What was the reported BE level? Higher concentrations make this explanation progressively less plausible."
[If contamination is raised:] "If there is concern about direct specimen contamination rather than systemic absorption, the specimen can be tested for PSA as a seminal marker."
"Is there anything specific about the clinical picture I can help you think through?"
Key distinction from semen-mediated transfer: The semen transfer scenario involves cocaine metabolites at extremely low concentrations (approximately 20-30 mcg per ejaculate) that are insufficient to produce a positive at standard cutoffs. Topical application involves the parent cocaine drug in direct mucosal contact, the same route confirmed to produce positive urine drug screens in medical literature.
| Scenario | Expected UDS | Basis |
|---|---|---|
| Direct applicator (cocaine applied to own penis) | Likely positive | Direct mucosal absorption; equivalent to TAC or nasal topical use; recreational dose may exceed medical doses substantially |
| Sexual partner (vaginal contact with cocaine-coated penis) | Plausible positive; concentration-dependent | Vaginal mucosa is an established absorption route (pharmaceutical drug delivery); no direct study exists, extrapolated from mucosal absorption pharmacology |
| Intact skin contact only (no mucosal surface) | Negative expected | Kavanagh 1992: passive aerosol and cutaneous exposure in medical staff produced no positive urine results |
Medical literature supporting mucosal absorption: Altieri et al. (1990) found 78% positive by EMIT and 83% by GC-MS after a single TAC topical anesthetic application, with results persisting 36-48 hours (PMID 2184707). Fitzmaurice et al. (1990) detected benzoylecgonine at 40 to >600 ng/mL in plasma within 20-40 minutes of TAC application in pediatric patients (PMID 2331095). The FDA nasal cocaine label explicitly warns of positive urine drug screens after clinical use.
Detection window: Benzoylecgonine (standard cocaine target) detectable approximately 24-72 hours after a single mucosal exposure. Cutoffs vary by test type and laboratory: POC cups commonly 300 ng/mL; clinical EIA/immunoassay typically 50-300 ng/mL (lab-dependent); LC-MS/MS confirmatory typically 5-150 ng/mL (lab-dependent); federal/forensic (DOT/SAMHSA) 150 ng/mL immunoassay, 100 ng/mL confirmatory. Confirm specific cutoffs with your laboratory before interpreting results.
[Clarify who is positive: the person who applied cocaine or their partner? Ask this before proceeding.]
"Before I address this, I want to make sure I understand the scenario. Which person tested positive: the one who applied the cocaine to their genitals, or their sexual partner?"
[If positive is in the direct cocaine applier:]
"In that case, a positive urine drug screen is pharmacologically expected. Cocaine absorbs through mucosal tissue. This is the same reason it is used as a topical anesthetic in ENT and wound care. Medical studies using topical cocaine wound anesthetics found 78 to 83% of patients test positive for benzoylecgonine after a single application, with results persisting up to 36 to 48 hours. The glans penis is mucosal tissue with similar absorption characteristics. This explanation is pharmacologically credible and should be documented."
[If positive is in the sexual partner:]
"That is a different scenario from the cocaine-in-semen question. Here we are talking about cocaine residue on the penis coming into direct contact with vaginal mucosa during penetration. The vaginal mucosa is a well-documented absorption route: it is used for pharmaceutical drug delivery. There is no published study specifically on this scenario, but the pharmacological basis is sound. The absorbed dose depends on residual cocaine concentration and contact duration. I would characterize this as plausible in principle but not confirmed by direct published data."
[If asked whether to accept or document the explanation:]
"For the direct applicator, this is a pharmacologically credible explanation that should be documented in the record. For the partner, document it as a reported exposure that is plausible but without direct supporting evidence. Either way, this is a substantially more credible claim than semen-mediated transfer."
| Use Pattern | Approximate Detection Window |
|---|---|
| Single use | 3–4 days |
| Moderate use (few times/week) | 5–7 days |
| Daily use | 10–15 days |
| Chronic heavy use | Up to 30+ days |
Hemp-derived CBD products: Regulated products must contain less than 0.3% delta-9-THC, but concentrated or impure products may produce a positive test, especially with heavy use.
| Isomer | Potency vs Delta-9 | Parent Half-Life | Clinical Profile |
|---|---|---|---|
| Delta-9-THC | 100% (reference) | 4–6 hrs | Classic intoxication; euphoria, impaired coordination; highest impairment risk |
| Delta-8-THC | ~50–70% | 4–8 hrs | Milder, calmer; less anxiety/paranoia; still impairing; widely sold in gas stations/vape shops |
| Delta-10-THC | ~20–40% | 4–8 hrs | More energizing/uplifting; lighter euphoria; "functional high" |
- Many retail Delta-8/Delta-10 products contain measurable Delta-9-THC due to poor manufacturing controls
- Chronic Delta-8 or Delta-10 use produces prolonged THC-COOH positivity identical to cannabis use
- Drug courts and monitoring programs typically treat all delta isomers identically
| Feature | Delta-9-THC | HHC (retail products) |
|---|---|---|
| Standard urine target | THC-COOH: detected on all panels | HHC-COOH-type metabolites: not included on most standard panels |
| LC-MS/MS confirmation | Confirmed as THC-COOH | Often missed or non-specifically reported; requires targeted HHC method |
Oral THC (edibles, capsules) produces proportionally more 11-OH-THC than inhaled cannabis due to first-pass hepatic metabolism. This explains why oral cannabis causes stronger, longer intoxication at equivalent doses and is a common cause of unexpected overdose with edibles.
11-OH-THC is not a routine urine confirmation target: standard panels report THC-COOH only.
Mechanism: Pyrethroids are sodium/chloride channel modulators that suppress GABA activity, producing a methamphetamine-like "rush."
Clinical risks: IV injection can cause multiorgan failure (lungs, heart, liver, kidneys). No specific antidote; treatment is supportive.
Geographic prevalence: Reported in KY, FL, NY, OH, TN, TX, WV.
Kratom (Mitragyna speciosa) contains two primary alkaloids detected by LC-MS/MS: mitragynine (the dominant alkaloid) and 7-hydroxymitragynine (7-OH). Standard immunoassay panels do not detect kratom. Specific LC-MS/MS testing reports both analytes quantitatively in ng/mL. Reference ranges have not been established for either analyte.
| Parameter | Mitragynine | 7-OH Mitragynine |
|---|---|---|
| Role in traditional kratom | Dominant alkaloid (~80-90% of alkaloid content) | Minor component (<2% of alkaloids in whole-leaf) |
| Mu-opioid receptor potency | Weak partial agonist; ceiling effect limits opioid-like risk | ~13x morphine; ~40x mitragynine (in vitro data only) |
| Source of 7-OH in urine | N/A | Metabolic conversion from mitragynine AND/OR direct ingestion of concentrated 7-OH product; cannot be distinguished by urine result alone |
| Reference ranges established? | No | No |
| Regulatory status | Not federally scheduled; DEA drug of concern | FDA recommending scheduling of concentrated products under CSA (July 2025) |
Product type is more clinically informative than analyte levels alone. Urine LC-MS/MS cannot distinguish traditional whole-leaf kratom use from use of a concentrated or synthetically derived 7-OH product. Ask about product form (whole-leaf powder, encapsulated supplement, extract, or 7-OH concentrate). Concentrated 7-OH products carry a risk profile comparable to conventional full mu-opioid agonists.
| Product Type | Typical Finding | Clinical Risk |
|---|---|---|
| Whole-leaf / powder / tea | Mitragynine predominant; low 7-OH if detected | Lower risk; partial agonist ceiling; withdrawal still possible |
| Kratom extract / full-spectrum | Higher mitragynine; 7-OH may be elevated | Higher potency; closer to conventional opioid risk profile |
| Concentrated / isolated 7-OH product | 7-OH disproportionately elevated relative to mitragynine | High risk: treat clinically as high-potency opioid; evaluate for co-ingestion |
| Product labeled as kratom (unknown) | Variable; any ratio possible | Cannot assume traditional leaf profile; obtain product details |
* Potency estimates (13x morphine, 40x mitragynine) are from in vitro and animal studies. No validated human morphine-equivalent conversion exists for 7-OH mitragynine. Do not use these figures for dose conversion. Deaths associated with kratom products have nearly all involved polysubstance use or co-contaminants; a positive result should prompt evaluation for other co-ingested substances.
[Gather: product type the patient reported using, symptom picture, other substances present, MAT status.]
"Thank you for calling. I understand you have a question about a kratom result, specifically about the mitragynine and 7-OH mitragynine levels."
"Mitragynine is the main alkaloid in traditional whole-leaf kratom. It acts as a relatively weak, partial opioid agonist; it has ceiling effects that limit the opioid-like risk compared to conventional opioids. The 7-OH mitragynine is far more potent, roughly 13 times morphine in animal studies, but in traditional whole-leaf kratom it makes up a very small fraction of the alkaloid content, so the overall risk profile is lower."
"The challenge is that we cannot tell from the urine result alone whether the 7-OH is coming from metabolic conversion of mitragynine (which happens normally) or from a concentrated 7-OH product the patient ingested directly. Those products (sometimes still labeled as kratom) carry a risk profile much closer to a conventional opioid."
"Right now, the most actionable clinical step is confirming what product the patient is actually using: whole-leaf powder, a standard supplement, or an extract or concentrated product. That product history is more informative than the numeric result alone at this point."
[If asked about morphine-equivalent dose:]
"There isn't a validated human dose-conversion equation for 7-OH mitragynine yet; the potency estimates are from in vitro and animal data. Use the result for risk stratification rather than dose conversion."
"Is there anything specific about the patient's clinical picture (symptoms, co-medications, or history) that I can help you think through?"
Nitrous oxide specifically: Half-life is approximately 5 minutes via exhalation. The standard diagnostic approach for chronic abuse is measuring B12 and homocysteine levels: nitrous oxide inactivates vitamin B12 and elevates homocysteine.
| Drug | Approximate OF Detection Window |
|---|---|
| Amphetamine / Methamphetamine | 1–3 days |
| Cocaine | 1–2 days |
| THC (cannabis) | 12–24 hours (current use indicator) |
| Opioids (morphine, codeine) | 1–2 days; cutoff 1 ng/mL |
| Buprenorphine | 2–3 days |
| Methadone | 1–2 days |
| Product | Route | OF Positivity Rate |
|---|---|---|
| Suboxone (sublingual) | Direct oral exposure | ~86% |
| Sublocade | SubQ depot injection | ~47% |
| Brixadi | SubQ depot injection | Comparable to or lower than Sublocade |
Breath mints: Generally not a concern for drug panels. They may transiently affect pH but do not produce false positives for drugs.
Example: cutoff is 20 ng/mL, raw result is 19.9 ng/mL. Rounding reports 20 (positive). Truncation correctly reports 19 (negative).
| Term | Definition | Clinical implication |
|---|---|---|
| Sensitivity | Ability to correctly identify true positives (drug present, test positive) | High sensitivity = fewer false negatives |
| Specificity | Ability to correctly identify true negatives (drug absent, test negative) | High specificity = fewer false positives |
LC-MS/MS: high sensitivity AND specificity: the definitive confirmatory standard.
Example: A 15 ng/mL codeine result against a 50 ng/mL cutoff is 3.3 times below the reporting threshold. This result is clinically and legally Negative.
| Parameter | Forensic (SAMHSA/DOT) | Clinical |
|---|---|---|
| Purpose | Employment, legal, government | Patient care, monitoring |
| pH acceptance | Up to 9.1 | Up to 9.5 |
| Chain of custody | Required | Not always required |
| Reporting | MRO review required | Direct physician interpretation |
- Alcohol-based hand sanitizers, mouthwashes
- Some foods (kombucha, very ripe fruit, certain breads)
- Certain medications
- Very remote or low-level exposure: EtS clears slightly faster than EtG and may fall below the cutoff before EtG does, producing an EtG-only result near the end of the detection window.
- Active infection: Certain bacterial organisms in the urinary tract can affect glucuronide readings; consider evaluating for UTI.
- Analytical issue: If the pattern is isolated, consider requesting a repeat specimen to confirm.
Three factors drive in-cup fermentation:
- Glucosuria (uncontrolled diabetes): Glucose spills into the urine and acts as a fermentation substrate for bacteria or yeast.
- Bacterial or yeast contamination in the collection cup.
- Prolonged processing delay: The longer the specimen sits at ambient temperature before processing, the more fermentation can occur.
General framework for evaluating a fermented food defense:
- EtG below 100 ng/mL: Dietary sources plausible; incidental exposure is possible.
- EtG 100 to 500 ng/mL: Possible; evaluate specific sources and consumption volume in context.
- EtG above 500 ng/mL: Dietary sources unlikely as the sole explanation; substantial cumulative intake would be required.
- EtG above 1,000 ng/mL: Fermented food sources insufficient to reliably produce this level.
A pattern of recurrent dilute specimens with intermittent EtG/EtS positives is more consistent with ongoing alcohol use combined with intentional system flushing than with incidental exposure. Documenting this pattern is important for building the clinical narrative.